Usability of a novel Hounsfield units measurement procedure to quantify intercorporal bone graft remodeling in patients after posterior lumbar interbody fusion: a case series

Abstract

Background

There is a lack of knowledge about the biological process of intercorporal bone graft remodeling after posterior lumbar interbody fusion surgery and whether this process is associated with changes in back pain and intercorporal fusion status. As an alternative to the commonly used but unreliable fusion criteria, Hounsfield units can be used to quantify biological activity and changes in bone mineral content. However, studies assessing Hounsfield units conducted to date do not provide sufficient details about how the bone grafts were segmented to measure the Hounsfield units to allow for replication, and did not assess individual patient trends in graft changes over time. Using the data of nine patients after posterior lumbar interbody fusion, a novel Hounsfield units measurement procedure was developed and used to explore its usability to quantify the bone graft remodeling process.

Case details

We report a case series of nine patients (six male, three female, mean age 64 years, all Caucasian) who all had computed tomography scans performed at 1 and 2 years after posterior lumbar interbody fusion surgery. Overall, seven out of the nine (78%) cases had a 3–41% increase in their bone grafts’ Hounsfield units between 1 and 2 years after surgery. The cases showed large interindividual variability in their Hounsfield units values over time, which coincided with varying levels of back pain and intercorporal fusion status.

Conclusion

The Hounsfield units measurement procedure used for this case series may be useful to quantify intercorporal bone graft remodeling in patients after posterior lumbar interbody fusion, and may be used as an adjunct diagnostic measure to monitor bone graft remodeling over time. Future research is warranted to explore how to interpret bone graft Hounsfield units-values and Hounsfield units trajectories in light of clinical variables or intercorporal fusion status.

Peer Review reports

Background

Lumbar spinal fusion is a surgical procedure that is widely being used for the treatment of a variety of spinal disorders and its incidence is rising [1, 2]. Surgical techniques to achieve fusion vary and consensus regarding a superior technique is lacking [3]. To promote interbody fusion the intervertebral disc is removed and replaced with an autologous or allogenic bone graft [4]. This nonvital bone graft has the ability to remodel into vital bone by going through different phases of bone regeneration [5]. This process is characterized by an increase in bone volume and bone quality and should ultimately lead to bony fusion of the two adjacent vertebrae. When this process fails, a nonunion or pseudarthrosis will occur. This important and well-known complication after spinal fusion affects 5–35% of patients and may lead to complaints of pain and a decrease in functional status [6]. Subsequent revision surgery for symptomatic pseudarthrosis is performed in up to 24% of patients [7].

In daily clinical practice, many surgeons and radiologists use plain radiography or computed tomography (CT) images to assess whether their patients’ intercorporal bone grafts are showing signs of increasing bone volume and quality and if the bone grafts are remodeling toward bony fusion or not. However, interpreting these images remains a serious challenge both in terms of how the bone graft is developing as well as ultimately judging whether vertebrae have fused or not [8,9,10]. A recent systematic review [11] has highlighted that assessment of intercorporal fusion is challenging, especially since there is no widely accepted definition of interbody fusion. This hinders confirmation of the real presence of solid bony fusion. Until a valid and practical method of assessing interbody fusion is available, quantifying the process of intercorporal bone graft remodeling toward fusion may be the most objective alternative for now.

Bone quality or bone mineral density (BMD) is traditionally measured using dual energy X-ray absorptiometry (DEXA) scans. However, this method is not appropriate for assessing patients’ bone graft BMD after spinal fusion surgery because the implanted instrumentation causes visual artefacts, which may affect BMD values [12]. Hounsfield units (HU), on the other hand, highly correlate with BMD [13] and can be measured from CT-images without interference from these artefacts. The use of HU as a means to assess biological activity and increased bone mineral content after lumbar fusion surgery was first described in the early 2000’s [14, 15]. More recent studies showed that HU can also be used to determine bone quality of vertebral bodies and around the bone fusion site [16, 17], further confirming the clinical utility of assessing bone quality using CT-attenuation in HU [18]. HU may also offer a promising method to objectively quantify intercorporal bone graft expansion or resorption over time, and ultimately assist in determining fusion status more objectively [19]. Unfortunately, the studies conducted to date did not provide sufficient details about how the bone grafts were segmented to measure the HU to allow for replication, and did not assess individual patient trends in bone graft changes over time. Therefore, we set out to develop a detailed and reproducible HU measurement procedure and to explore its usability to quantify the bone graft remodeling process in nine patients after posterior lumbar interbody fusion (PLIF).

Case details

Available data from nine Caucasian patients that consented to participate in a previous trial [20] was collected for this case series. The characteristics of these cases are summarized in Table 1, cases were treated with single-level PLIF surgery using autologous bone graft that was impacted behind a polyetheretherketone (PEEK) cage to promote interbody fusion. No intra- or postoperative complications nor other adverse events were reported for any of the cases.

Table 1 Case characteristics

All cases had a 1-year and 2-year follow-up CT-scan using one of three different scanners (Philips iCT 256, Philips Brilliance 40, Siemens Emotion 16) using different kernels (HNP/Myelo and Trauma protocols), kilovoltage peaks (ranging between 100 and 140) and filter types (A or D), which is known to influence HU outcomes [21]. Because no phantom model was used during the original trial we deemed it necessary to calculate a correction factor on the basis of the accessible anatomical structures also imaged on the cases’ CT scans. Details about this procedure and all the cases’ raw data can be found here: https://data.mendeley.com/datasets/632dx4vb96/5.

Hounsfield unit measurement procedure

For the HU measurement procedure we used the dynamic oblique multiplanar reformatting option of the Sectra IDS7 software program to position each cases’ CT-scan with the lower endplate of the vertebrae above, and the upper endplate below the grafted area aligned as horizontally as possible. Each image was saved in a 1 mm slice thickness setting and then viewed using the Sectra UniView medical image viewer with the dorsal structures displayed on the right. Subsequently, the observer scrolled through the different sagittal CT-slices to judge in which of these the largest and most hyperintense region of bone graft was visible. This slice was chosen as the primary slice of interest (pSOI) and used as the starting point for all subsequent HU measurements. To ensure that the exact same pSOI could be localized for the intraobserver reliability measurements, as well as for the two-year follow-up CT-scans, the observer scrolled through the sagittal CT-slices from the pSOI until the first available cage marker was visible. This cage marker was used as the reference slice to re-identify the pSOI. Second, using the software program’s freehand drawing option, in the pSOI a region of interest (ROI) was demarcated around the visible area of the cases’ bone graft, resulting in an automatically calculated HU value (see Fig. 1).

Fig. 1
figure 1

Measuring the intercorporal bone graft’s Hounsfield units of case 8. Freehand drawn regions of interest including the intercorporal bone graft at 1 (left scan) and 2 (right scan) years after surgery. The (0.5 mm) cage markers, which acted as reference points beyond which the regions of interest was demarcated (posteriorly), are clearly visible on both scans. In this case the (raw, uncorrected) HU increased from 488 HU to 536 HU over time

Care was taken that the ROI did not include the adjacent lumbar body endplates nor the cage. The latter was ensured by consistently demarcating the ROI posterior to the reference cage marker in all CT-slices used. This exact same procedure was repeated using five adjacent CT-slices to the cases’ right and four to the cases’ left of the pSOI, respectively. Subsequently, the resulting HU values from all CT-slices were summed and divided by the number of each cases’ available CT-slices (typically ten) to obtain a standardized estimate (mean) of the bone graft’s HU both for the 1 year and 2 year CT-scan slices. The 2 year HU outcomes were multiplied with the calculated correction factor. The intraobserver reliability of the measurement procedure was additionally established using 174 CT-slices of the nine cases using their 1-year, and another 174 CT-slices using their 2-year CT scans both measured with a 2 week interval. The resulting intraclass coefficient (ICC) was 0.93 [95% confidence interval (CI) 0.91–0.95, single-rater, consistency, two-way mixed-effects model].

The cases’ changes in bone graft HU values from 1-year to 2-years after surgery are presented in Fig. 2A and their percentage change over time are visualized in Fig. 2B. The correction factors of the cases ranged between 0.66 and 1.49. Overall, there was a wide variety in the cases’ individual postsurgical characteristics, which did not always correspond with each other. For example, one male patient’s (case 8) intercorporal bone graft showed an increase from 545 HU to 623 HU over time (Fig. 1). This patient’s process of bone remodeling coincided with a clear decrease in back pain levels as assessed using a visual analogue scale (VAS) and a decrease in perceived disability related to back pain as assessed using the Oswestry disability index (ODI). Both at 1 and 2 years after surgery, the two vertebrae of this case were subjectively judged as being fused. One female patient (case 2) also showed an increase in HU, with decreases in VAS back pain scores (from 87 to 8) and ODI scores (from 42 to 6) over time, but both at 1 and 2 years after surgery the two vertebrae of this case were judged as not being fused. Overall, seven cases showed an increase in their HU over time, representing a change between 3% and 41%. Two cases showed a negative change of −14% and −31% in their overall HU. One of these was a male patient (case 6) who kept reporting back pain and who showed unchanged ODI scores over time despite the orthopedic surgeon judging his two vertebrae as being fused at 2 years after surgery. The development of the HU of the other male patient (case 1) also suggested that the bone quality was decreasing over time. Nevertheless, this cases’ back pain decreased, his health state as assessed using the European Quality of Life 5 Dimensions Health Questionnaire (EQ-5D-5L) increased notably and his vertebrae were judged as being fused.

Fig. 2
figure 2

Change in the cases’ bone grafts Hounsfield units over time using internal phantomless calibration. The change in Hounsfield units from 1- to 2-years postoperatively (A) and the percentage change, with the first year being the starting point (B). Each line represents one individual case

Discussion

Open surgical exploration is the gold standard procedure to confirm lumbar fusion status. However, this procedure is too impractical and considered unethical [11]. As such, quantifying the bone graft remodeling process itself may currently be the only appropriate and reliable method that can inform the orthopedic surgeon about whether their patients’ intercorporal bone grafts are remodeling toward fusion.

Using the data of nine patients after PLIF allowed us to explore the usability of a novel Hounsfield units measurement procedure to quantify intercorporal bone graft remodeling. The excellent intraobserver reliability of the measurement procedure (ICC 0.93) suggested that the measurement procedure was highly reproducible on an individual observer’s basis, and this gave us the confidence to subsequently explore the cases’ bone grafts HU over time. Our cases showed expected large interindividual variability in for example, their HU values, back pain levels, and health status. Overall seven out of the nine (78%) cases had a 3–41% increase in their bone grafts’ HU between 1 and 2 years after surgery.

To our knowledge, only two previous studies have investigated changes in HU over time as a proxy measure of the bone graft remodeling process after spondylodesis. One of these studies [14], published almost 20 years ago, revealed an increase of 23% in HU of the autografts inside fusion cages at 1 year after surgery. Subsequently, a 6% decrease in HU was detected at 2 years. In that study the HU values from all patients were pooled without reporting on baseline differences that affect bone quality, which makes it susceptible to confounding as this could influence the bone grafts’ HU values. In addition, the use of early generation CT scanners may also have influenced the HU values in terms of artefact influence and bias from limitations in scanner calibration. A second study also reported increasing HU-values in individual patients’ intercorporal bone grafts after spondylodesis up to 2 years after spinal fusion surgery [15]. However, in this study single measurements of the participants’ bone graft HU-values were used for pooled trend analysis and details about the scanning procedures were lacking. Therefore, our protocol included a calibration correction factor when different types of CT-scanners were used.

Given the positive association between HU and BMD [13], the increases in HU values of seven of the cases suggest that their bone was still remodeling up to 2 years after surgery, and their increasing bone density may have been a sign of bone graft proliferation toward bony fusion. Conversely, in the two cases (22%) that showed a decrease in their HU, this may have been a sign of bone resorption, which could ultimately lead to fusion failure and pseudoarthrosis. This latter finding is in line with the results of research showing that 5–35% of patients develop pseudoarthrosis after spine fusion surgery [22]. However, for now associations between the downward, upward, or stable trajectories in the bone grafts’ HU and other clinical characteristics, such as back pain, health status and the surgeon’s judgements of fusion status remain elusive. This is illustrated by different cases showing either decreased (for example, cases 2 and 8) or unchanged (for example, cases 3 and 9) back pain levels together with increases in their bone grafts’ HU and health status. Similarly, cases showing clear increases in their bone grafts’ HU were judged as fused (for example case 8) or not fused (for example, case 4), while the two cases (1 and 6) who showed clear decreases in their bone grafts’ HU were both judged as being fused.

The patients’ data presented in this case series have shown that it is currently still unclear how to interpret bone graft HU-values and HU trajectories in light of clinical variables, such as the patients’ level of back pain, perceived disability related to back pain and health status.. Interpretation is also seriously hindered by the unreliability of current fusion judgement criteria [11]. Future research, using much larger patient samples and assessing repeated CT-scans in the first year after PLIF surgery, is needed to gain better insight into the development of HU values over time and thus the process of bone graft remodeling. This will also assist in obtaining more precise and reliable estimates of associations between the different clinical outcomes. Despite this limitation, monitoring the patients’ HU-values over time obtained using our measurement procedure may be a valuable adjunct clinical outcome alongside the unreliable criteria currently used to judge whether this has ultimately led to intercorporal fusion. For example, imagine a patient who is reporting persistent back pain after PLIF surgery and about whom the orthopedic surgeon is unsure about the need for revision surgery given that the patient’s vertebrae are (subjectively) judged as being fused (for example, our cases 3 and 9). Revisiting these patient’s available old and new CT-scans and using these to establish the bone grafts’ HU over time could assist in gauging whether intercorporal bone was indeed expanding, not changing or resorbing over time. In case of clear increases in the patients’ bone graft’s HU, bone remodeling toward fusion would more likely compared with when HU were not changing or even decreasing. In case of the latter situations, the results of additional diagnostics, such as flexion–extension radiographs [23] or the facet fluid sign [24], could be used to further assist in clinical decision-making.

Conclusion

The Hounsfield units measurement procedure used for this case series may be useful to quantify intercorporal bone graft remodeling in patients after PLIF, and be used as an adjunct diagnostic measure to monitor bone graft remodeling over time. Future research is warranted to explore HU trajectories in light of clinical variables and intercorporal fusion status.

Availability of data and materials

The datasets used and analyzed during the current study are available and freely accessible at https://data.mendeley.com/datasets/632dx4vb96/5.

Abbreviations

BMD:

Bone mineral density

CT:

Computed tomography

HU:

Hounsfield units

ICC:

Intraclass coefficient

PLIF:

Posterior lumbar interbody fusion

References

  1. Malik K, Nelson A. Overview of low back pain disorders. In: Essentials of pain medicine. Elsevier; 2018. p. 193- 206.e2. https://doi.org/10.1016/B978-0-323-40196-8.00024-3.

    Chapter  Google Scholar 

  2. Martin B, Mirza S, Spina N, Spiker W, Lawrence B, Brodke D. Trends in lumbar fusion procedure rates and associated hospital costs for degenerative spinal diseases in the United States, 2004–2015. Spine. 2018;44:1.

    Google Scholar 

  3. Mobbs RJ, Phan K, Malham G, Seex K, Rao PJ. Lumbar interbody fusion: techniques, indications and comparison of interbody fusion options including PLIF, TLIF, MI-TLIF, OLIF/ATP LLIF and ALIF. J Spine Surg. 2015;1(1):2–18.

    PubMed  PubMed Central  Google Scholar 

  4. Cole CD, McCall TD, Schmidt MH, Dailey AT. Comparison of low back fusion techniques: transforaminal lumbar interbody fusion (TLIF) or posterior lumbar interbody fusion (PLIF) approaches. Curr Rev Musculoskelet Med. 2009;2(2):118–26.

    Article  PubMed  PubMed Central  Google Scholar 

  5. Gupta A, Kukkar N, Sharif K, Main BJ, Albers CE, El-Amin Iii SF. Bone graft substitutes for spine fusion: a brief review. World J Orthop. 2015;6(6):449–56.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Chun DS, Baker KC, Hsu WK. Lumbar pseudarthrosis: a review of current diagnosis and treatment. Neurosurg Focus. 2015;39(4):E10.

    Article  PubMed  Google Scholar 

  7. Martin BI, Mirza SK, Comstock BA, Gray DT, Kreuter W, Deyo RA. Reoperation rates following lumbar spine surgery and the influence of spinal fusion procedures. Spine. 2007;32(3):382–7.

    Article  PubMed  Google Scholar 

  8. Brantigan JW. Pseudarthrosis rate after allograft posterior lumbar interbody fusion with pedicle screw and plate fixation. Spine (Phila Pa 1976). 1994;19(11):1271–9; discussion 80.

    Article  CAS  PubMed  Google Scholar 

  9. Christensen FB, Laursen M, Gelineck J, Eiskjaer SP, Thomsen K, Bünger CE. Interobserver and intraobserver agreement of radiograph interpretation with and without pedicle screw implants: the need for a detailed classification system in posterolateral spinal fusion. Spine (Phila Pa 1976). 2001;26(5):538–43; discussion 43-4.

    Article  CAS  PubMed  Google Scholar 

  10. Diedrich O, Perlick L, Schmitt O, Kraft CN. Radiographic characteristics on conventional radiographs after posterior lumbar interbody fusion: comparative study between radiotranslucent and radiopaque cages. Clin Spine Surg. 2001;14(6):522–32.

    CAS  Google Scholar 

  11. Duits AAA, van Urk PR, Lehr AM, Nutzinger D, Reijnders MRL, Weinans H, et al. Radiologic assessment of interbody fusion: a systematic review on the use, reliability, and accuracy of current fusion criteria. JBJS Rev. 2024. https://doi.org/10.2106/JBJS.RVW.23.00065.

    Article  PubMed  Google Scholar 

  12. Spiro AJ, Hoang TD, Shakir MKM. Artifacts affecting dual-energy X-ray absorptiometry measurements. AACE Clin Case Rep. 2019;5(4):e263–6.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Pinto EM, Neves JR, Teixeira A, Frada R, Atilano P, Oliveira F, et al. Efficacy of Hounsfield units measured by lumbar computer tomography on bone density assessment: a systematic review. Spine. 2022;47(9):702–10.

    Article  PubMed  Google Scholar 

  14. Burkus JK, Dorchak JD, Sanders DL. Radiographic assessment of interbody fusion using recombinant human bone morphogenetic protein type 2. Spine. 2003;28(4):372–7.

    Article  PubMed  Google Scholar 

  15. Spruit M, Meijers H, Obradov M, Anderson PG. CT density measurement of bone graft within an intervertebral lumbar cage: increase of Hounsfield units as an indicator for increasing bone mineral content. J Spin Disord Tech. 2004;17(3):232–5.

    Article  CAS  Google Scholar 

  16. De Stefano F, Elarjani T, Warner T, Lopez J, Shah S, Basil GW, et al. Hounsfield unit as a predictor of adjacent-level disease in lumbar interbody fusion surgery. Neurosurgery. 2022;91(1):146–9.

    Article  PubMed  Google Scholar 

  17. Song J, Taghavi CE, Hsu DW, Song K-J, Song J-H, Lee K-B. Radiological changes in anterior cervical discectomy and fusion with cage and plate construct: the significance of the anterior spur formation sign. Spine. 2012;37(4):272–9.

    Article  PubMed  Google Scholar 

  18. Zaidi Q, Danisa OA, Cheng W. Measurement techniques and utility of Hounsfield unit values for assessment of bone quality prior to spinal instrumentation: a review of current literature. Spine (Phila Pa 1976). 2019;44(4):E239-e44.

    Article  Google Scholar 

  19. Brantigan JW, Steffee AD. Carbon fiber implant to aid interbody lumbar fusion: 1-year clinical results in the first 26 patients. Lumbar Fusion and Stabilization. Cham: Springer; 1993. p. 379–95.

    Google Scholar 

  20. Lehr AM, Oner FC, Delawi D, Stellato RK, Hoebink EA, Kempen DHR, et al. Efficacy of a standalone microporous ceramic versus autograft in instrumented posterolateral spinal fusion: a multicenter, randomized, intrapatient controlled, noninferiority trial. Spine (Phila Pa 1976). 2020;45(14):944–51.

    Article  PubMed  Google Scholar 

  21. Free J, Eggermont F, Derikx L, van Leeuwen R, van der Linden Y, Jansen W, et al. The effect of different CT scanners, scan parameters and scanning setup on Hounsfield units and calibrated bone density: a phantom study. Biomed Phys Eng Express. 2018;4(5): 055013.

    Article  Google Scholar 

  22. Brodsky AE, Kovalsky ES, Khalil MA. Correlation of radiologic assessment of lumbar spine fusions with surgical exploration. Spine (Phila Pa 1976). 1991;16(6 Suppl):S261–5.

    Article  CAS  PubMed  Google Scholar 

  23. Choudhri TF, Mummaneni PV, Dhall SS, Eck JC, Groff MW, Ghogawala Z, et al. Guideline update for the performance of fusion procedures for degenerative disease of the lumbar spine. Part 4: radiographic assessment of fusion status. J Neurosurg Spine SPI. 2014;21(1):23–30.

    Article  Google Scholar 

  24. Aggarwal A, Garg K. Lumbar facet fluid-does it correlate with dynamic instability in degenerative spondylolisthesis? A systematic review and meta-analysis. World Neurosurg. 2021;149:53–63.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the staff of the Radiology department at Rijnstate Hospital for their invaluable cooperation during this research.

Funding

There was no funding associated with this research.

Author information

Authors and Affiliations

Authors

Contributions

JFHR designed the study, collected, analyzed, and interpreted the data, and drafted the manuscript. JLCvS was the originator of the research concept, assisted in interpreting the data, and was a major contributor in writing the manuscript. MCK contributed to writing the manuscript and interpreting the data. MJvG advised about the technical aspects and interpretation of radiological images and contributed to writing the manuscript. LDdJ provided project supervision, assisted in interpreting the data, and was a major contributor in writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Joey F. H. Reijmer.

Ethics declarations

Ethics approval and consent to participate

The original fusion trial (https://clinicaltrials.gov/ct2/show/NCT01982045) was approved by the medical ethics review committee of the University Medical Center Utrecht and local institutional review boards, and was in accordance with the principles of the Declaration of Helsinki (version October 2008) and the Medical Research Involving Human Subjects Act. Approval of this study was also granted by Rijnstate Hospital’s institutional research ethics review board (study number 2020-1663). All cases provided written informed consent to participate in the trial.

Consent for publication

Written informed consent was obtained from the patient for for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Reijmer, J.F.H., van Susante, J.L.C., Kruijt, M.C. et al. Usability of a novel Hounsfield units measurement procedure to quantify intercorporal bone graft remodeling in patients after posterior lumbar interbody fusion: a case series. J Med Case Reports 18, 337 (2024). https://doi.org/10.1186/s13256-024-04655-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13256-024-04655-4

Keywords