Fig. 7

From: Combined flow-based imaging assessment of optimal cardiac resynchronization therapy pacing vector: a case report

Fig. 7

Quantitative analysis (to compare with the results of analysis represented in Fig. 3), by echo-particle imaging velocimetry of the orientation angle (φ) of the global hemodynamic forces exchanged between blood and surrounding tissues during acute study (post-cardiac resynchronization therapy 6-month follow-up). a Two-dimensional transthoracic echocardiogram high-temporal resolution contrast echocardiography is performed by three-chamber apical view approach (see Additional file 10: Video S10 interventricular − 50 ms SONOVUE 6-month follow-up). b The polar histogram showed with setting interventricular delay − 50 ms the most aligned intraventricular forces that were partially dominated by the longitudinal path of pressure gradient (lower φ, value 43.1°). CRT cardiac resynchronization therapy, F.U. follow-up, PIV particle imaging velocimetry, TTE transthoracic echocardiogram, VV interventricular